GH3220T is an ultra-low-power and ultra-high-precision heart rate sensor that integrates multi-channel PPG, single-lead ECG, and dual-channel temperature sensing circuits. It monitors health indicators such as heart rate (HR), heart rate variability (HRV), peripheral capillary oxygen saturation (SpO2), and pulse wave translation time (PWTT), and body & skin temperature. GH3220T is ideal for high-end smart watches, smart wristbands, ECG patches, professional sports gear, and medical devices.
Ultra-Low Power Consumption
Average current consumption in heart rate detection mode: 59 µA
Average current consumption in SpO2 detection mode: 230 µA
Ultra-high Performance
24-bit highly accurate ADC
PPG dynamic range: up to 110 dB
Up to 256 µA configurable Background Light Cancellation (BGC)
Up to 256 µA configurable DC offset cancellation
AC Background Light Rejection (BGR) (50 Hz): 60 dB
DC Background Light Rejection (BGR): 80 dB
ECG Common-Mode Rejection Ratio (CMRR): 100 dB
ECG input impedance > 2 GΩ,support dry electrode
Input noise meet YY-0885/IEC60601-2-4
Temperature Sensing Module
2 NTC temperature sampling channels and 1 ref resistance sampling channel
Temperature sensing range and deviation: 25 °C ≤ Temp. < 45 °C: ±0.1 °C
Light Transmitter Block
Two built-in LED drivers: each supports 4 LEDs, total 8 LEDs
Each LED driver with 8-bit resolution and a maximum drive current of 200 mA
Supports 8 LEDs concurrently for SpO2 or multi-wavelength and multi-position heart rate detection
Light Receiver Block
Four AFEs that support synchronous sampling (when the temperature sensing channels are not in use)
Up to 8 programmable sample time-slots; up to 32 effective channels for data acquisition
2 PDs in differential connection or 4 PDs in single-ended connection
Build-in Functions
Automatic dimming
Automatic wrist (wearing) detection
Right leg drive (RLD) single-lead ECG sensing circuit
Internal FIFO
800 * 4 bytes, caching 800 samples
Communication Interface
SPI: up to 12 MHz
IIC: up to 1 MHz
Critical Electrical Characteristics
AVDD: 2.0 V – 3.6 V
VDDIO: 1.62 V – 3.6 V
Operating temperature: -30 °C – 85 °C
Package
WLCSP Package
Size: 2.605 mm * 2.905 mm * 0.468 mm, 0.4 mm pitch, 42 balls
Typical Applications Supported
Heart Rate
SpO2
ECG
PWTT
HRV
Wearing detection
Temperature measurement
Part Number | GH3220T | |
---|---|---|
PPG TX | LED interface | 8 |
Max LED Driving Current | 200 mA * 2 | |
PPG RX | AFE channels | 4 |
PD interface | 5 Single/2 Diff | |
Max SNR | 110 dB | |
ECG | ECG channels | 1 |
Input Impedance | 2 GΩ | |
RTI noise | 1.36 µVrms | |
Power Consumption | 590 µA@500SPS | |
Temperature | NTC interface | 3 |
Temperature measurement accuracy | ±0.1°C@25°C~ 45°C | |
Interface | IIC/SPI | |
ADC | 24-bit | |
FIFO | 3.2k bytes | |
Package Type | WLCSP-42 | |
Chip Size(mm) | 2.6 * 2.9 * 0.46 | |
Algorithm | HR, SpO2, HRV, ECG, Wear Detect |
Name | Rights Description | Version | Date |
---|---|---|---|
GH3220T_Datasheet | Register Users Only | Rev.1.3 | 2022-01-06 |
GHealth_User_Manual | Nda Users Only | Rev.1.2 | 2021-08-17 |
GH(M)3x2x EVK Hardware User Guide | Nda Users Only | Rev.1.0 | 2020-08-04 |
GH3220T_Schematic_BOM | Nda Users Only | Rev.1.1 | 2023-07-13 |
GH3220T_Schematic | Register Users Only | Rev.1.1 | 2022-02-14 |
Open WeChat, use "Scan" to follow.